FX Series FAQs

FXseries-001

Question
While using FX, the message "Port opening failed" appeared.
Answer
Turn off the instrument and reset it (turn it on while holding down the Enter key).
Order
001

FXseries-007

Question
Is there any way to make the alphabet capitalized?
Answer
Switch characters with the [α] key, press and hold the [Shift] key to switch lowercase and uppercase letters.
Order
007

FXseries-008

Question
Data has been erased, but can it be recovered?
Answer
Not possible.
Order
008

FXseries-009

Question
The password was not set, but the screen came out.
Answer
It cannot be used until the correct password is entered. To unlock the device, it is necessary to send the instrument to the purchasing company.
Order
009

FXseries-010

Question
FX series prism constants, How to change the prism constants?
Answer
Change with "Target" of "EDM setting" in "Settings".
Order
010

FXseries-011

Question
Distance resolution: Is there a 0.1mm setting? (FX series)
Answer
It is not used in the FX series. It is used in SX, DX, and iX.
Order
011

FXseries-012

Question
Can the battery BDC72 be used with the FX series instead of BDC70?
Answer
Can be used. The charger for BDC72 is CDC77.
Order
012

FXseries-013

Question
Why does the FX series shift about 1' when measuring just above zenith?
Answer
Near the zenith and directly below, the tilt correction is cut off, so the tilt of the instrument will appear as it is. Align the center with the electronic bubble tube on the screen.
Order
013